Technical Specification
BodyGuard 323MR™
Piston Mechanism

Infusion Modes
Continuous

Dose

Concentration Units
mg/ml, U/ml
Dose Units
mg/kg/min or hr
mcg/kg/min or hr
U/kg/min or hr
mU/kg/min or hr
Non-Weight Dose

Concentration Unit
mg/ml, U/ml
Dose Units
mg/min or hr
mcg/min or hr
U/min or hr
mU/min or hr
Delay Before Start

Time/Volume Mode

Flow Rate Accuracy
± 5 %
Flow Rate Range
0.1-99.9 ml/h; 0.1ml/h Steps
100-1200 ml/h; 1ml/h Step
Adjustable Maximum Rate

Maximum Bolus Rate
1200 ml/h
Programmable Infusion Time
Up to 100 hrs
Volume to be Infused
0.1-99.9 ml; 0.1 ml Steps
100-9999 ml; 1 ml Steps
Maximum Bolus Volume
100 ml
KVO Rate
0 (OFF) - 5 ml/h
Resume on Restart

Ultrasonic Air Sensor
Single Bubble
0.1-1 ml
Accumulated Air-in-Line

Occlusion Pressure
5, 7.5 or 10 psi
Pressure Sensor Mechanism

Post-Occlusion Bolus Reduction

Pump Alarms/Alerts
Air-in-Line

Down Occlusion

Low Battery

End Program

End Battery

Pump Unattended

Door Open

Lock Mode

Missing Key

Display
Size (W/H)
6x2 cm
Resolution (W/H)
122x32 pixels
Power Supply
100 – 240V AC, 47-63 Hz (5.0A)

Battery
Rechargeable Li-Polymer 7.4V 1800 mAh.
Life at 125ml/h:

Up to 15hrs
Connectivity (Optional)
RS232 Serial Port

802.11 WiFi

BodyComm™ Software (Optional)

Remote Monitoring System (Optional)

EMR Interface

Protection Against Electrical Shock
Pump/Bracket
Type CF (internally powered
External Power Supply
Type CF, Class I
Moisture Protection
IPX1
Dimensions; pump only (H/W/D)
112 x 89 x 40 mm
Weight (with standard battery)
390g
* Specification may vary depending on software version.
** Manufacturer reserves the right to change the specifications without prior notification.