Technical Specification

T34™ 

Linear Syringe Driver Mechanism

Infusion Modes

     Continuous

          Lock ON

          Lock OFF

          Rate Mode

     PCA

     Time mode

Syringe Size*

2 – 50 ml

Syringe Brands

All brands programmable

Drug Volume Auto-Detect

Flow Rate Accuracy

± 2 %

Flow Rate Range

0.1-99.9 ml/h; 0.1 ml/h Steps

100-650 ml/h; 1 ml/h Step

Programmable Infusion Time

10 mins – 100 hrs

Adjustable Maximum Bolus Volume

Max Syrine Travel

67 mm

KVO Rate

0 (OFF) - 2 ml/h

Resume on Restart

Occlusion Pressure

100–1500mmHg;

10mmHg Step

Maximum Force

50N

Pump Alarms/Alerts

     Syringe not Recognized

     Occlusion

     End Program

     Near End/Nearly Complete

     End Syringe Travel

     Low Battery

     End Battery

     Pump Paused

     Lock Mode

     Syringe Displaced

     Check Syringe Placement

Display

     Size (W/H)

6x2 cm

     Resolution (W/H)

122x32 pixels

Power Supply
100 – 240V AC, 50/60 Hz (10W / 0.3A)

     Rechargeable Li-Polymer

     7.4V 1800 mAh. Life at 25ml/h:

     Recharge Time (Empty to Full) :

Battery

     Disposable Alkaline 9V.
     Life at 125ml/h:

Up to 14 hrs

Connectivity (Optional)

BodyComm™ Software (Optional)

Remote Monitoring System (Optional)

PDMS Interface

Classification

Type CF

Moisture Protection

IPX3

Dimensions; pump only (H/W/D)

167 x 66 x 39 mm

Weight (with battery)

277g

* Max volume infused for single infusion varies between syringe brands.
** Specification may vary depending on software version.
*** Manufacturer reserves the right to change the specifications without prior notification.

  1. DATA SGP