Technical Specification
T34™
Linear Syringe Driver Mechanism

Infusion Modes
Continuous
Lock ON

Lock OFF

Rate Mode

PCA
Time mode
Syringe Size*
2 – 50 ml
Syringe Brands
All brands programmable
Drug Volume Auto-Detect

Flow Rate Accuracy
± 2 %
Flow Rate Range
0.1-99.9 ml/h; 0.1 ml/h Steps
100-650 ml/h; 1 ml/h Step
Programmable Infusion Time
10 mins – 100 hrs
Adjustable Maximum Bolus Volume

Max Syrine Travel
67 mm
KVO Rate
0 (OFF) - 2 ml/h
Resume on Restart

Occlusion Pressure
100–1500mmHg;
10mmHg Step
Maximum Force
50N
Pump Alarms/Alerts
Syringe not Recognized

Occlusion

End Program

Near End/Nearly Complete

End Syringe Travel

Low Battery

End Battery

Pump Paused

Lock Mode

Syringe Displaced

Check Syringe Placement

Display
Size (W/H)
6x2 cm
Resolution (W/H)
122x32 pixels
Power Supply
100 – 240V AC, 50/60 Hz (10W / 0.3A)
Rechargeable Li-Polymer
7.4V 1800 mAh. Life at 25ml/h:
Recharge Time (Empty to Full) :
Battery
Disposable Alkaline 9V.
Life at 125ml/h:
Up to 14 hrs

Connectivity (Optional)

BodyComm™ Software (Optional)

Remote Monitoring System (Optional)
PDMS Interface
Classification
Type CF
Moisture Protection
IPX3
Dimensions; pump only (H/W/D)
167 x 66 x 39 mm
Weight (with battery)
277g
* Max volume infused for single infusion varies between syringe brands.
** Specification may vary depending on software version.
*** Manufacturer reserves the right to change the specifications without prior notification.