Technical Specification
BodyGuard 575 Color Vision™
Piston Mechanism
Linear Syringe Driver Mechanism
Infusion Modes
Basal Only
Basal & Bolus
Bolus Only
Auto Bolus
PiAB
Drug Library
Number of Drug Protocols
26 Protocols
Syringe Size*
Syringe Brands
Concentration Units
mg/mcg
Flow Rate Accuracy
±5 %
Basal Rate Range
0.1-100 ml/h; 0.1 ml/h Steps
Adjustable Maximum Rate
Maximum Bolus Rate
1200 ml/h
Volume to be Infused
0.1-99.9 ml; 0.1 ml Steps
100-9999 ml; 1 ml Steps
Max Syringe Travel
KVO Rate
0(OFF) - 5 ml/h
Resume on Restart
Ultrasonic Air Sensor
Single Bubble
0.1 – 2 ml
Accumulated Air-in-Line
Occlusion Pressure
250 - 520 mmHg; 10mmHg Steps
PCA
PCEA
Regional
Pressure Sensor Mechanism
Actual Pressure Display & Graph
Post-Occlusion Bolus Reduction
Pump Alarms/Alerts
Air/Up Occlusion
Down Occlusion
End Program
Near End
End Syringe Travel
Low Battery
End Battery
Door Open
Pump Unattended
Lock Mode
Missing Key
Syringe Displaced
Check Program
Check Syringe Placement
Display
Size (W/H)
3.7x4.9 cm
Power Supply
100 – 240V AC, 50/60 Hz (10W / 0.3A)
Resolution (W/H)
240x320 pixels
Battery
Rechargeable Li-Polymer 7.4V 1800 mAh.
Life at 125ml/h:
Recharge Time (Empty to Full):
Up to 17 hrs
6hrs
Disposable Alkaline 9V (x2).
Life at 125ml/h:
Up to 12 hrs
Extended-life Rechargeable Li-Polymer 7.4 V
3600 mAh. Life at 125ml/h:
Recharge Time (Empty to Full)
Up to 34 hrs
10hrs
Connectivity (Optional)
RS232 Serial Port
802.11 WiFi
BodyComm™ Software (Optional)
Remote Monitoring System (Optional)
EMR Interface
Classification
Type CF, Class II
Moisture Protection
IPX3
Dimensions; pump only (H/W/D)
112 x 89 x 40 mm
Weight (with standard battery)
390g
* Max volume infused for single infusion varies between syringe brands.
** Specification may vary depending on software version.
*** Manufacturer reserves the right to change the specifications without prior notification.